Support Nextteq® CBA Kit

Nextteq® CBA Kit Support


Product Overview


Nextteq®'s new CBA kit including Nextteq® regulator and flow meter system.

Our new CBA kit now includes the ability to test for oxygen!  In addition, a new Nextteq® regulator and flow meter system is included to work with the Nextteq® CBA tubes. 


Quality is key and our new regulator and flow meter system is no exception.  The flow meter comes with a calibration chart prepared by comparing the flow meter to a calibrator.  Use of the chart will improve the accuracy of the set flows, especially at the lower end of the flow meter scale.  Flow meters tend to be more accurate a higher flow rates than they are at lower flow rates.  The calibration chart will allow for more accurate readings for all of the tubes included in the CBA kit.


Testing is made simple!  Once you attach the regulator to an air bottle or you adapt it to an air supply manifold, follow the 3 steps below for all tubes except O₂:

  • Break the tube tips and insert the tube with the arrow pointing down
  • Set the flow rate for the first tube
  • At the end of the allotted time, remove the tube from the regulator, and read the tube immediately!


For an accurate reading of O₂, you will be using a syringe kit with included instructions.  The syringe is used to precisely monitor the volume of 50cc sample air.


Support Documents


CBA Regular Kit Manual

CBA Regular Kit Manual

Nextteq® Detector Tube Guide

Nextteq Detector Tube Guide

Additional Support


For more information, contact Nextteq Customer Service Team at 813-249-5888 or toll free at 877-312-2333.

For technical questions, ask to speak to Dr. Duffy, VP of Technical Development / Industrial Hygiene Manager.